Contents:
1) Introduction
2) Irradiator Description
3) Irradiator Sources
4) Irradiator Shielding
5) Survey Meter
6) Operating Procedures
7 Emergency Procedures
8) Required Form
|
|
In case of an emergency
involving the Gammacell Irradiator, Contact the Radiation Safety Office Immediately.
The following conditions are considered emergency or
abnormal conditions warranting immediate notification of the Radiation Safety Office:
|
 |
Fire in the area. |
 |
An exposure reading on the survey instrument
exceeding 2000 uR/hr |
 |
Failure of the sources to return to the fully
retracted position. |
Emergencies
During Operation
| 1) Press "Stop" on the keypad. |
2) If the sources do not retract as
indicated by a display reading of: "Return Motor Timeout
Fault" |
You Should:
 |
Turn the key switch to "Off". |
 |
Vacate the area. |
 |
Contact Radiation Safety immediately. |
Emergency Contacts
| MCG Radiation Safety: |
1-9826 |
| Dr. Anatolij Horuzsko |
1-8736 |
| Dr. Patricia Schoenlein |
1-6281 |
In case of fire:
| 1) Contact MCG Public Safety, 1-2911. |